![]() |
|

| X-Rite 301 Densitometer. |
![]() |
Performance Features: The 301 Densitometer provides highly repeatable and accurate measurements of black and
white film densities up to 5.0 D – more subtle than can be seen by the naked eye.An optional RS232 output allows you to connect to your current computer. Easy to Use: Operation is as simple as “push and read,” and the large LED numerals can be read easily in bright or dim light. Internal memory and the null button allow the operator to make comparative density measurements across a piece of film. Functional Performance Specifications
Measuring Range: 0-5.0D with 2 and 3mm apertures / 0-4.0D with 1mm aperture
Accuracy: ±.02D Repeatability: ±.01D Operating Temperature Range: 10° - 40°C 50° - 104°F Warm-Up Time: 60 seconds Scale Factor (slope) Stability: ±1% per 6 months Null Drift: ± .03D max. / ± .01D typical Physical Dimensions: 5.25” H (13.3cm) X 10.25” W (26.0cm) X 15.0” L (38.0cm) Weight: 8.5lbs (3.9kg) Additional down-loadable documents on this product... |
|
| X-Rite 331 Portable Transmission Densitometer. |
![]() |
Performance Features: Though the 331 Densitometer is portable and compact, it has the same accuracy and
repeatability as larger countertop units, measuring densities up to 4.0 D.Easy to Use: The built-in light table eliminates the need for an external light source and easily accommodates film up to eleven inches wide. The 331 has push-button zeroing, an on/off switch to prevent battery drainage, a certified step wedge, a low-battery indicator, and instrument carrying case. Functional Performance Specifications
Measuring Range: 0-3.5D (1mm) 0-4.0D (2mm) Accuracy: ±.02D Repeatability: ±.01D Operational Temperature Range: +50°F - +104°F (+10°C - +40°C) Storage Temperature Range: -22°F - +122F (-30°C - +50°C) Warm-Up Time: None Scale Factor (slope) Stability: ±1% per 6 months Zero Stability per 8 hours: ±.02D (.01 typical) Measuring Area: 1 & 2 mm Measuring Length: 5.5 inches Power Supply: 4 rechargeable AA Ni-Cad batteries and charger Physical Dimensions: 2.0” H (5.08cm) X 2.9” W (7.46cm) X 7.0” L (17.78cm) Weight: 1.5lbs (680g) Instrument / 2.7lbs (1.2kg) Shipping. Additional down-loadable documents on this product... |
|
| X-Rite 396 Sensitometer. |
![]() |
Sensitometers imprint a standard set of exposures on a photographic or x-ray film sample. When the film sample is
developed, a densitometer is used to read the optical density of the exposures and chart a profile against a known set
of standards. This profile alerts you to fluctuations in processing conditions and allows you to take corrective action.Small and convenient, the 396 Dual-Color Sensitometer provides single-sided exposures and produces repeatable 21 step exposures on film. Set the battery operated 396 to the proper exposure, place the film inside, then firmly press down on the cover. When you hear the beep, the exposure is complete. DIN Specifications: Designed and manufactured to comply with the calibration requirements described in DIN V 6868-55. |
| Functional Performance Specifications
Exposure Stability: ±.02 log exposure per year Unit to Unit Repeatability: ±.02 log exposure Temperature Sensitivity: ±.02 log exposure from 15° C; (59°F) to 30°C (86°F) Power Requirement: 9 volt alkaline battery (included) approx. 10,000 exposures/battery Light Modulation: 21-step Wedge, 0.15D per step Blue Color Peak Wavelength: 455nm ± 10nm Green Color Peak Wavelength: 512nm ± 10nm Arm-Up Time: None Exposure Time: 0.1 second typical regulated by light accumulation system Recycle Time 2 seconds Physical Dimensions: 2.3" H (5.84cm) X 3.75" W (9.50cm) X 7.0" L (17.78cm) Weight: 0.9 lbs. (410g) Additional down-loadable documents on this product... |
|
![]() |
|