| Fluke Biomedical Stock Number: 06-524-3000 |
![]() ![]() ![]() |
IMAGE INTENSIFIER ION CHAMBER Designed to measure diagnostic x-rays. This 100 cc low profile ion chamber is designed specifically for measuring exposure rate at the input phosphor of fluoroscopic image intensifier tubes. Its extremely thin envelope enclosure allows for the most accurate results possible! Its unique size and shape allow it to be inserted into the spot film tray of typical image intensifier systems. A 21" long detachable handle allows for easy insertion and removal. When used with the RAD-CHECK MICRO-R, it is possible to measure exposure rates as low as 10 micro Roentgens per second. SPECIFICATIONS: Volume: 100 cc Nominal Sensitivity: 30 nC/R @ M75 Rate: 10µR/s Exposure: 100µR in 10 seconds Max. Exposure Rate: Min. 90% collection at 20R/Sec Total Attenuation: 0.275 (mm Al) equivalent @ 30 KeV Operating Voltage: 200 to 300 volts DC Handle Length: 21" (53.34 cm) Cable Length: 10' (3 m). Cable Termination, BNC/Banana Window Thickness: .110" (2.79 mm) Housing Material: Acrylic Housing Color: Clear Housing Dimensions: 5.5" diameter (14.0 cm diameter) Surface Area: 100 cm², 4.5" diameter Thickness: 0.720" (18.3 mm) Weight: 30 oz (340 g) |
![]() |
|