| Progeny Linear IV Collimator System |
![]() | The Linear IV Collimator System is a microprocessor controlled, fully automatic
unit designed to provide accurate collimation for overtable radiographic,
fluoroscopic, and special procedure applications. The Linear IV provides
automatic collimation even in applications where both the x-ray source and the
receptor are movable and not limited to fixed SID's. Features: |
| Radiographic Ratings: 50Kvp Rated - Meets all requirements of 21 CFR
Sub-Chapter J, and IEC 601-1-3. Ray Proofing: Less than 50mR/Hr at one meter from focus spot of x-ray tube measured at 150kV at 4mA leakage factor. Projection Lamp: Quick change, pre-aligned FDS pre-focused Quartz Haloagen lamp operates on 24 VAC. Lamp times out after approximately 25 seconds and may be reset at any time. Quick turn-off features allow for limited lamp use. Light output > 160 Lux at 40 inches. Tube Mounting: Mounted by 4 bolts at 90 with 3.625" bolt line center. Swivel mount allows 360 rotation with 90 detents. Dimensions: 8.10" x 6.99" x 12.03" - 205.7mm x 177.5mm x 305.6mm - Weight = 19.8 lbs. (9 kg). Color: Light grey (standard) Other colors available upon request. Electrical: 120/240 VAC, 50/60 Hz, 8 Amps. Certification: Certified to UL-2601-1 CAN/CSA-C22.2 No. 601-1-M-90, and IEC 601-2-32. Reference UL File No. E181750. CE Approved. |
![]() |
|