| Doppler String Phantom. |
| Specifications...
Flow Simulation Speeds: 10 to 200 centimeters per second, bi-directional. Speed Drift: Crystal-locked to 20 parts per million (0.002%). Accuracy: +/- 1% of stated speed. Pulsatile Waveforms: 16 pre-programmed and optional customer specified. Waveforms Included: Adult common carotid, stenotic carotid, femoral, aortic. Fetal middle cerebral artery, renal artery, umbilical artery. Pediatric descending thoracic artery, patent ductus arteriosus. Test waveforms: Sine waves with peak speeds of 100, 150, and 200 cm/second. Triangle waves with peak speeds of 100, 150, and 200 cm/second. Stepped ramp wave with stops at 0, 20, 40, 60, 80, and 100 cm/sec. Waveform Resolution: Each waveform simulation contains 1000 points of resolution, or speed adjustments, enabling extremely complex simulation. Computer Interface: Industry standard RS-232 interface built-in for future enhancements and remote control. Very useful for automated quality control in a manufacturing environment. Fluids Used in Tank: Plain tap water (velocity 1480 meters per second at 20 degrees C) or velocity corrected water/glycol solution giving 1540 meters per second at 20 degrees C. Phantom adjusts itself for either fluid. Physical Specifications: 120 Volts AC, 50 Watts. Total weight in travel case: 22 lbs. (10Kg). Travel case dimensions: 17" x 17" 10" (25 x 19.5 x 14 cm). Tank dimensions: 10" x 12" x 9" deep (25 cm x 30 cm x 23 cm ). | ![]() |
|
![]() |
|