Pro-RF Geometry Collimation/Beam Alignment Test Tools, with Cone and/or Cylinder-
This device readily indicates a 1% or 2% misalignment at a 1m focal-film distance (FFD), but it may be used at any FFD.
It consists of a flat 20 x 25 cm plate with a 14 x 18 cm pattern etched on its surface. It can also be used to check
fluoroscopy alignment and collimation.
Click for details: 
A steel ball is mounted in the center of a disc at each end of the 15 cm high clear plastic cylinder. When the balls are
positioned over one another and at a right angle to the film, their images will appear as one if the central ray is
truly perpendicular to the film. The approximate degree of improper angulation can also be determined.
TECHNICAL DATA: (can be modified to customer specifications):
Dimensions: 200 x 250mm
Copper plate with engraved 5mm scale
150 x 60 mm cylinder with a steel ball embedded in the center of its bases – beam alignment
cone for the perpendicular X-Ray beam control in the range of 0° ÷ 1.5° (option, instead of the cylinder – for simpler
measurements)
PRODUCT FEATURES:
Complies with: IEC 61223-3-3, CE certified, Manual provides detailed guidelines for carrying out each test, results
assessment and registration.