Pacific Northwest X-Ray Inc.

       TNT 12000 X-Ray Test Device.       


Fluke Biomedical, Inc. Technical Data

The TNT 12000 X-Ray Test Device is the newest and most versatile instrument available for measuring key x-ray imaging parameters. It sets up in seconds and measures kVp, dose, dose rate, time, and half value layer (HVL) in a single exposure. A totally-wireless ZigBee® interface enables quick and easy setup and the wireless detector can be used with the companion wireless display or a laptop computer.

X-ray imaging QA, calibration, and maintenance in today’s demanding digital environment require very high productivity and compliance with local and regional regulations. The TNT 12000 delivers high productivity through 100 % wireless connectivity to either a display, which instantly responds with all values in a single exposure, or a laptop, where measured values are displayed and categorized in organized templates. Instant HVL with just a single exposure further enhances productivity.

Key features

  • 100% ZigBee wireless operation between detector and hand-held display or laptop
  • Compact hand-held design for maximum portability and ease-of-use
  • Simple user interface with minimum menu routines means setup in seconds
  • Fluke Biomedical ruggedness provides reliable operation. It’s tough!
  • 40 kHz sampling rate to ensure accuracy with the most difficult applications
  • Global support network delivering prompt service and peace-of-mind to Fluke Biomedical customers worldwide

  • Product positioning
    The TNT 12000 takes over where other technology leaves off. Featuring all-in-one-exposure measurements and ZigBee wireless communication combined with the rugged, reliable, and accurate design that is a Fluke trademark, the TNT 12000 is truly a new breed of noninvasive x-ray test tool.

    TNT 12000 is ideal for use by OEM factory and field service engineers, independent service organizations, physicists, biomedical and clinical engineers, and local and regional field inspectors of x-ray imaging equipment. Its small, lightweight design enhances portability and 100 % wireless operation ensures setup can be accomplished in seconds. Because the TNT 12000 measures all parameters with every exposure, there is no need for complicated menu selection, further enhancing user productivity. TNT 12000 always


    defaults to the last use when powered on, so when used often for repetitive procedures it is truly a one-button (power-on) solution. The TNT 12000 has the expanded functionality needed for modern applications and can be managed with minimum keystrokes. Users can identify and select custom measurement protocols and save them for future use. Full test automation and documentation software is available for TNT 12000, creating the advantage of accurate, repeatable testing processes. Ansur Test Automation Software is only available from Fluke Biomedical.

    Primary end benefit
    Productivity and accuracy are central goals when performing any maintenance or QA process on diagnostic imaging equipment because image system uptime is critical to patient care objectives. TNT 12000 is small, portable, and wireless. It is easily transported to the imaging room, set up in seconds, and all results are instantly available on the hand-held display or on the user’s laptop. The latter application places measured values into templates that are standard with the Excel software provided with TNT 12000. Because the TNT 12000 Excel package is a spreadsheet, users can customize their templates and create their own reports to send to others. Accuracy, reproducibility, and reliability are also critically important and Fluke Biomedical is the industry standard customers count on to deliver uncompromised performance.

    Technical specifications

    Physical specifications
    Display: 320 × 240 Color LCD
    Size (WxDxH)
    Display: 15.2 cm x 11.4 cm x 4.45 cm (6 in × 4.5 in × 1.75 in)
    Detector: 15.2 cm x 11.4 cm x 4.45 cm (6 in × 4.5 in × 1.75 in)
    Weight
    Display: 0.422 kg (0.93 lb)
    Detector: 0.68 kg (1.5 lb)

    Electrical specifications
    Battery
    Battery type: Lithium-ion, 3.7 V, 4000 mAh
    Battery charge time: Approx. 5 hr
    Battery discharge time: Approx. 8 hr
    Battery cutoff voltage: 2.75 V
    AC adapter






    Input voltage: 100 V ac to 240 V ac
    Input frequency: 50/60 Hz
    Input current: 0.5 A (rms)
    Output voltage: 6 V dc

    Environmental specifications
    Operating temperature: 0 °C to 35 °C (32 ºF to 95 ºF)
    Storage temperature: -35 °C to 50 °C (-31 ºF to 122 ºF)
    Operating humidity: 20 % to 80 % RH (non-condensing)

    kVp measurements
    Units: kVp Average (average of peaks during a specified interval)
    kVp Max: (highest peak during a specified interval) PPV (peak practical voltage)

    Ranges
    Radio/Fluoro modes: 40 kV to 150 kV
    Mammo modes:
    Mo/Mo: 22 kV to 35 kV (standard calibration)
    Rh/Rh: 25 kV to 49 kV (optional calibration)
    Mo/Rh: 22 kV to 40 kV (optional calibration)
    Mo/Al: 22 kV to 49 kV (optional calibration)
    Rh/Al: 25 kV to 49 kV (optional calibration)
    W/Rh: (optional calibration)
    W/Ag: (optional calibration)

    Resolution: 0.1 kV
    Accuracy: (including calibration uncertainty)
    Radio/Fluoro modes: ± 5 %
    Mammo modes: ± 5 %
    Reproducibility: ± 1 % (std of 5 readings)
    Filtration correction range
    Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent
    Mammo mode: 0 mm Al to 0.4 mm Al added filtration



    Dose/Exposure measurements
    Units: Roentgens, grays
    Range: 10 mR to 999 R, 0.1 mGy to 999 Gy
    Resolution: 1 μR 0.01 μGy
    Accuracy: ± 5 % (including calibration uncertainty)
    Reproducibility: ± 1 % (std of five readings)
    Filtration correction range
    Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent
    Mammo mode: 0 mm Al to 0.4 mm Al added filtration
    kV correction ranges
    Radio/Fluoro modes: 40 kV to 150 kV
    Mammo mode: Mo/Mo: 22 kV to 35 kV

    Dose/Exposure rate measurements
    Units: Roentgens or grays per hour, minute, second, pulse
    Range
    50 mR to 19.9 R/s
    0.5 mGy to 199 mGy/s
    2 μR/pulse to 333 mR/pulse (@ 60 PPS)
    0.02 μGy/pulse to 3.33 mGy/ pulse (@ 60 PPS)
    Resolution
    1 μR/s
    0.01 μGy/s
    0.02 μR/pulse (@ 60 PPS)
    0.2 nGy/pulse (@ 60 PPS)
    Accuracy
    ± 5 % (including calibration uncertainty)
    Filtration correction range
    Radio/Fluoro modes: 1 mm Al to 10 mm Al or equivalent
    Mammo mode: 0 mm Al to 0.4 mm Al added filtration
    kV correction range
    Radio/Fluoro modes: 40 kV to 150 kV
    Mammo mode: Mo/Mo: 22 kV to 35 kV







    Exposure time—radiographic modes
    Units: Milliseconds, pulses
    Range: Milliseconds: 10 ms to 999 ms; Pulses: 1 pulse to 999 pulses
    Resolution: Milliseconds: 0.1 ms; Pulses: 1 pulse
    Accuracy: Milliseconds: 1 % or 0.5 ms; Pulses: ± 1 pulse
    Reproducibility: Milliseconds: 1 % or 0.5 ms; Pulses: ± 1 pulse

    Elapsed time—fluoro modes
    Range: 10 sec to 999 sec
    Resolution: 0.1 second
    Accuracy: 1 % or 0.5 sec

    Average pulse rate—pulsed fluoro
    Range: 1 pps to 999 pps (pulses per second)
    Resolution:1 pps
    Accuracy: 1 pps

    Average pulse width— pulsed fluoro
    Range: 10 ms to 999 ms
    Resolution: 0.1 ms
    Accuracy: 1 % or 0.5 ms

    HVL
    Range
    Radio/Fluoro modes: 1.2 mm Al to 10 mm Al (equivalent)
    Mammo mode: 0.2 mm Al to 0.4 mm Al (equivalent)
    Resolution
    Radio/Fluoro modes: 0.1 mm Al (equivalent)
    Mammo mode: 0.01 mm Al (equivalent)
    Accuracy
    Radio/Fluoro modes: ± 10 % or 0.2 mm Al (equivalent)
    Mammo mode: ± 5 % or 0.05 mm Al (equivalent)


    Stock #
    Kit/Model Description
    Included Accessories in this Kit.
    Your Price
    3335763
    TNT 12000 Wireless Detector + Display
    3345931 - TNT 12000WD Wireless Detector
    3345954 - TNT 12000D Wireless Display.
    3346048 - AC Adapter(s)/Charger(s)
    3346027 - USB Cable, Type A to Mini B
    3346030 - USB Cable, Mini A to Mini B
    3341333 - ZigBee USB Dongle
    3346011 - Customized Carrying Case
    3586667 - CD-ROM, DoseMate/TNT/mAs
    3557133 - Quick Start Reference Guide
    $12,993.75
    3335774
    TNT 12000WD Wireless Detector
    3345931 - TNT 12000WD Wireless Detector
    3346048 - AC Adapter(s)/Charger(s)
    3346027 - USB Cable, Type A to Mini B
    3341333 - ZigBee USB Dongle
    3346011 - Customized Carrying Case
    3586667 - CD-ROM, DoseMate/TNT/mAs
    3557133 - Quick Start Reference Guide
    $11,043.75
    3335795
    TNT 12000D Wireless Display
    3345954 - TNT 12000D Wireless Display
    3346048 - AC Adapter(s)/Charger(s)
    3346027 - USB Cable, Type A to Mini B
    3346030 - USB Cable, Mini A to Mini B
    3341333 - ZigBee USB Dongle
    3586667 - CD-ROM, DoseMate/TNT/mAs
    $2,990.00
    3586807
    TNT 12000WD Wireless Detector with
    mAs + DoseMate + Display
    3586570 - TNT 12000WD Wireless Detector with mAs
    3561890 - TNT 12000 DoseMate Dosimeter
    3345954 - TNT 12000D Wireless Display
    3548014 - AC Power Adapter3346027 - Cable, Type A to Mini B USB
    3346030 - Cable, Mini A to Mini B USB
    3586537 - Test Stand
    3586644 - Triax Cable, Male to Male, BNC, 6 ft
    3341333 - ZigBee USB Dongle
    3264115 - HVL Filter Set
    3549414 - Multi-Prong Adapter
    3264091 - Adapter Stem
    3555490 - Test Lead Kit, 1 Set of Test Leads
    1918780 - mA Cable Assembly, BNC, male to female
    $18,193.75
    3586818
    TNT 12000WD Wireless Detector with
    mAs + DoseMate (no display)
    3586570 - TNT 12000WD Wireless Detector with mAs
    3561890 - TNT 12000 DoseMate Dosimeter
    3346048 - AC Adapter(s)/Charger(s) with multi-prong adapters
    3346027 - Cable, Type A to Mini B USB
    3346030 - Cable, Mini A to Mini B USB
    3586537 - Test Stand
    3586644 - Triax Cable, Male toMale, BNC, 6 ft
    3341333 - ZigBee USB Dongle
    3264115 - HVL Filter Set
    3264091 - Adapter Stem
    3555490 - Test Lead Kit, 1 Set of Test Leads
    1918780 - mA Cable Assembly, BNC, male to female
    3586555 - mA Shunt Assembly
    1938315 - Connector, BNC (female) to Binding Posts
    $16,243.75


    Stock #
    Optional Accessories
    Your Price
    35035
    mA/mAs Meter
    $1,243.75
    3337356
    TNT 12000 Ansur Test Automation
    Software Plug-in
    $853.28







    Cards We Accept!
    For order methods and policies, please click here.


    E-Mail Sales
    Telephone
    All Lines open 8-5 M-F (Pacific Time)
    Toll-Free (US): 800-827-XRAY (800-827-9729)
    Toll-Free (US): 866-PNW-XRAY (866-769-9729)
    Normal: 1-503-667-3000
    Download our order form
    and then...Fax Your Order
    24 Hours a day!
    1-503-666-8855

    P.O. Box 625 * Gresham, OR  97030 * U.S.A.

    This web site is our catalog, no printed catalog is available. - Please keep checking back, changes to this site are made daily.
    Every effort is made on our part to keep all posted prices up to date, however we can not guarantee the accuracy due to notification delays by the manufacturers.
    Purchase Orders accepted only upon approval of credit - Min. $500 opening order for new open CHARGE accounts.
    All other orders (Cash, Credit Card, etc) subject only to a $30 product minimum order!



    ©1997-2012 Pacific Northwest X-Ray Inc. - Sales & Marketing Division - All Rights Reserved